Mostrar el registro sencillo del ítem

Simultaneous determination of glycated haemoglobin, a long term biomarker of diabetes mellitus, and total haemoglobin by isotope dilution and HPLC-ICP-MS

dc.contributor.authorCastillo Busto, María Estela del 
dc.contributor.authorMontes Bayón, María 
dc.contributor.authorAñón Álvarez, Elena
dc.contributor.authorSanz Medel, Alfredo 
dc.date.accessioned2013-01-30T10:04:02Z
dc.date.available2013-01-30T10:04:02Z
dc.date.issued2008
dc.identifier.citationJournal of Analytical Atomic Spectrometry, 23, p. 758-764 (2008); doi:10.1039/b718008jspa
dc.identifier.issn0267-9477
dc.identifier.urihttp://hdl.handle.net/10651/7640
dc.description.abstractThe measurement of glycated haemoglobin (particularly HbA1c) serves as a powerful indicator for the evaluation and treatment of patients with diabetes mellitus. The separation of the glycated form of this protein (present about 3–4% in non-diabetic patients) is achieved by cation exchange chromatography, employing a Mono S column. The detection is carried out by measuring the Fe contained in the haemo-group of the protein, using an inductively coupled plasma mass spectrometer (ICP-MS) as the Fe selective detector. Using this experimental set-up, speciation of glycated and non-glycated haemoglobin has been investigated and detection limits around 1 μg mL−1 of protein have been obtained. Furthermore, accurate determination of the Fe naturally present in glycated and non-glycated haemoglobin peaks can be conducted by post-column addition of a solution containing isotopically labeled iron (57Fe, 92.4%) and applying isotope dilution calculations. The proposed method has been validated by determining glycated and non-glycated Hb (physiological levels about 10–13 g dL−1) in reference standards commonly used to calibrate clinical analyzers and also in a reference material (RM 405). Additionally, the ICP-MS based quantification method has been applied to the analysis of 11 blood specimens of patients diagnosed with diabetes. The relative results for HbA1c are compared to those obtained for the same samples by an external clinical laboratory. A good agreement in all cases was found. Finally, the potential of the complementary use of electrospray mass spectrometry (ESI-MS) is shown to study peak purity, identity and purity of the HPLC separated fractions before on-line ICP-MS detection.spa
dc.format.extentp. 758-764spa
dc.language.isoeng
dc.publisherRoyal Society of Chemistry
dc.relation.ispartofJournal of Analytical Atomic Spectrometryspa
dc.sourceWOKspa
dc.subjectBiochemical, Blood, Inductively Coupled Plasma/Ms, Lc/Ms, Medical, Quantitative Organic Analyses.spa
dc.titleSimultaneous determination of glycated haemoglobin, a long term biomarker of diabetes mellitus, and total haemoglobin by isotope dilution and HPLC-ICP-MSeng
dc.typejournal article
dc.identifier.local786spa
dc.identifier.doi10.1039/b718008j
dc.relation.publisherversionhttp://dx.doi.org/10.1039/b718008jspa