Evaluation of pulsed radiofrequency glow discharge time-of-flight mass spectrometry for precious metal determination in lead fire assay buttons

dc.contributor.authorCompernolle, Sien
dc.contributor.authorPisonero Castro, Jorge 
dc.contributor.authorBordel García, Nerea 
dc.contributor.authorWambeke, Dorine
dc.contributor.authorRaedt, Ine de
dc.contributor.authorKimpe, Kristof
dc.contributor.authorSanz Medel, Alfredo 
dc.contributor.authorVanhaecke, Frank
dc.date.accessioned2013-01-30T09:57:53Z
dc.date.available2013-01-30T09:57:53Z
dc.date.issued2011
dc.identifier.citationAnalytica Chimica Acta, 701(2), p. 129-133 (2011); doi:10.1016/j.aca.2011.05.043spa
dc.identifier.issn0003-2670
dc.identifier.urihttp://hdl.handle.net/10651/6320
dc.description.abstractIn this paper, an exploration of the capabilities and limitations of pulsed radiofrequency glow discharge time-of-flight mass spectrometry (GD-TOFMS) for the determination of the precious metals Ag, Au, Pd, Pt and Rh in lead buttons obtained by Pb fire assay is reported on. Since the matrix consists almost entirely of lead (>99%), the occurrence of doubly charged Pb (Pb2+) ions can hinder accurate determination of Rh. This problem was counteracted by relying on the time-resolved formation of different ion types over the pulse period of the glow discharge, which allows discrimination against the Pb2+ ions. The formation of ArCu+ ions as a result of the use of a copper anode was assessed to pose no threat to the accuracy of the results obtained for the set of samples analyzed as its contribution to the total signal at m/z = 103 could be adequately corrected for. The method developed was evaluated in terms of accuracy and precision using a set of Pb button standards with analyte concentrations between 5 and 100 μg g−1. For the purpose of validation, a 10 μg g−1 standard was considered as a sample. Overall, an acceptable accuracy was obtained with a bias of <5% between the experimental results and the corresponding reference values, except for Au, for which a larger deviation occurred. Precision values (repeatability) of typically <5% relative standard deviation (RSD) (for N = 3) were obtained and the limits of detection (LODs) vary from ∼0.020 μg g−1 for Ag to ∼0.080 μg g−1 for Pt.spa
dc.format.extentp. 129-133spa
dc.language.isoeng
dc.relation.ispartofAnalytica Chimica Actaspa
dc.rights(c) Analytica Chimica Acta
dc.sourceWOKspa
dc.titleEvaluation of pulsed radiofrequency glow discharge time-of-flight mass spectrometry for precious metal determination in lead fire assay buttonsspa
dc.typejournal article
dc.identifier.local20111073spa
dc.identifier.doi10.1016/j.aca.2011.05.043