Mostrar el registro sencillo del ítem

Application of a force sensor to improve the reliability of measurement with articulated arm coordinate measuring machines

dc.contributor.authorGonzález Madruga, Daniel 
dc.contributor.authorCuesta González, Eduardo 
dc.contributor.authorBarreiro García, Joaquín
dc.contributor.authorFernández-Abia, Ana Isabel
dc.date.accessioned2014-04-04T08:14:22Z
dc.date.available2014-04-04T08:14:22Z
dc.date.issued2013
dc.identifier.citationSensors (Switzerland), 13(8), p. 10430-10448 (2013); doi:10.3390/s130810430
dc.identifier.issn1424-8220
dc.identifier.urihttp://hdl.handle.net/10651/25235
dc.description.abstractA study of the operator contact force influence on the performance of Articulated Arm Coordinate Measuring Machines (AACMMs) is presented in this paper. After developing a sensor capable of measuring the contact force applied by an operator, a ring gauge has been used to analyse the relationship between the contact force and diameter and form errors measured with the AACMM. As a result, contact force has been proved as one of the main factors influencing the AACMM performance. A probe deflection model based on the Finite Element Method (FEM) has been also proposed in order to obtain the AACMM probe deflection caused by contact force. This allows measurement correction by comparing them with reference values, specifically, a ring gauge. Experimental test results show a significant measurement improvement that minimizes diameter error. Finally, an uncertainty evaluation for the contact force sensor and AACMM measurements with and without probe deflection model has been carried out in order to validate the ability of the sensor and the methodology followed
dc.description.sponsorshipMinistry of Finance and Competitiveness of Spain for the support with the research project ‘Quality Assurance and Knowledge Modelling Applied to Portable Coordinate Measuring Systems’ (ref. DPI2012-22642-C02-01), to the Instituto Universitario de Tecnología de Asturias (IUTA) for the support with the project ref. SV-12-GIJON-1 and to the Manufacturing Engineering Society (SIF) and the R2TAF for the mobility grant that have funded the work
dc.format.extentp. 10430-10448
dc.language.isoeng
dc.publisherMDPI
dc.relation.ispartofSensors, 13(8)
dc.rightsCC Reconocimiento - No comercial - Sin obras derivadas 4.0 Internacional
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.titleApplication of a force sensor to improve the reliability of measurement with articulated arm coordinate measuring machinesen
dc.typejournal article
dc.identifier.local20141096
dc.identifier.doi10.3390/s130810430
dc.relation.projectIDMEC/DPI2012-22642-C02-01
dc.relation.projectIDUniversidad de Oviedo-IUTA/SV-12-GIJON-1
dc.relation.projectIDSIF/R2TAF
dc.relation.publisherversionhttp://dx.doi.org/10.3390/s130810430
dc.rights.accessRightsopen access